• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC - GALWAY FLEXTOME® CUTTING BALLOON®; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC - GALWAY FLEXTOME® CUTTING BALLOON®; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS Back to Search Results
Model Number H749CBM3400150
Medical Device Problem Code Detachment Of Device Component (1104)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 03/05/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
It was reported that balloon detachment occurred.A 4.0 x 15 flextome cutting balloon was selected for use.During preparation, outside patient's body, the balloon came off with blue sheath covering the balloon.There were no patient complications reported.
 
Additional Manufacturer Narrative
Device evaluated by manufacturer: the device was returned for analysis.The balloon material was noted to have completed detached from the shaft of the device and was returned still within the blue balloon protective sheath.Multiple kinks were also noted along the length of the hypotube.Balloon material was noted to have completely detached from the shaft of the device and was returned still held within the blue balloon protective sheath.A visual and microscopic examination identified that the balloon material had completely detached from the shaft of the device.It detached from the device approximately 8mm proximal to the proximal edge of the proximal markerband.The balloon material was returned within the blue balloon protective sheath.The investigator removed the balloon material from the sheath.No issues were noted with the balloon material that could have contributed to the complaint incident.Visual and microscopic investigation identified no damage or any issues with the blades, tip or markerbands of the device that could have contributed to the complaint incident.All blades were present and fully bonded to the balloon material.Visual and tactile examination of the hypotube noted multiple kinks along its length.This type of damage is consistent with excessive force being applied to the delivery system.The balloon protector was received in position over the balloon.Due to the condition of the returned device it was not possible to apply a vacuum; however, the investigator removed the balloon protector sheath and slight resistance was encountered.The lack of vacuum may have resulted in the encountered resistance.The returned balloon protector inner dimension was verified at 0.0425 inch using a pin gauge and was within the specified range.No issues were identified that could have contributed to the complaint incident.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The investigation conclusion is handling damage as the complaint was caused by handling of the device or portion of the device without direct patient contact.(b)(4).
 
Event or Problem Description
It was reported that balloon detachment occurred.A 4.0 x 15 flextome cutting balloon was selected for use.During preparation, outside patient's body, the balloon came off with blue sheath covering the balloon.There were no patient complications reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLEXTOME® CUTTING BALLOON®
Common Device Name
CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS
Manufacturer (Section D)
BOSTON SCIENTIFIC - GALWAY
MDR Report Key7389309
Report Number2134265-2018-02385
Device Sequence Number679929
Product Code NWX
UDI-Device Identifier08714729808480
UDI-Public08714729808480
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P950020
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date (Section B) 03/06/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/09/2020
Device Model NumberH749CBM3400150
Device Catalogue NumberCBM340015
Device Lot Number20510124
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/30/2018
Initial Date Received by Manufacturer 03/06/2018
Supplement Date Received by Manufacturer04/19/2018
Initial Report FDA Received Date04/02/2018
Supplement Report FDA Received Date05/15/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-