| Model Number |
6207-D1 |
| Medical Device Problem Codes |
Break (1069); Material Integrity Problem (2978)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
02/07/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Evaluation summary: the introducer was returned and analyzed.The mechanical operation of the catheter shaft was kinked/buckled.The mechanical operation of the catheter was damaged.Visual analysis of the lead indicated damage during use.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that the lead introducer wings had broken off too early with resulted in the physician needing scissors to complete the "peal-away" procedure.The introducer was removed.No patient complications have been reported as a result of this event.
|
| |
|