| Catalog Number |
326800 |
| Medical Device Problem Code |
Manufacturing, Packaging or Shipping Problem (2975)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
03/08/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This device is same and/or similar to a device approved for distribution in the u.S.
|
| |
|
Event or Problem Description
|
|
It was reported that the packaging of a jelco® viavalve® safety iv catheter was white with no orange strip.No injury was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Customer returned 2 (5 up strips) of jelco® viavalve® safety iv catheters missing the 14ga "orange color bar" for evaluation.Upon review of the two strips, they were confirmed to be (cavity #'s 1-5 and 6-0) represented one full machine cycle.In the packaging operation, 10 blister packages are produced simultaneously.The investigation did confirm that the product failed to meet the expected performance as a result of a manufacturing related issue.
|
| |
|
Search Alerts/Recalls
|