Catalog Number 326800
Medical Device Problem Code Manufacturing, Packaging or Shipping Problem (2975)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 03/08/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).This device is same and/or similar to a device approved for distribution in the us.
 
Event or Problem Description
It was reported that the packaging of a jelco® viavalve® safety iv catheter was white with no orange strip.No injury was reported.
 
Additional Manufacturer Narrative
The complainant returned two "5 up strips" that were missing the 14ga "orange color bar" for evaluation.Upon reviewing the two strips, they were confirmed to be cavity #'s 1-5 and 6-0, which represented one full machine cycle.In the packaging operation, 10 blister packages are produced simultaneously.The investigation did confirm that the product failed to meet the expected performance as a result of a manufacturing related issue.Device history record was reviewed.The product was packaged on ffs #11 on the 22nd of april 2017.No discrepancies or abnormalities relevant to the complaint.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JELCO® VIAVALVE® SAFETY IV CATHETER
Common Device Name
CATHETER,INTRAVASCULAR,THERAPEUTIC,SHORT-TERM LESS THAN 30 DAYS
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
6000 nathan lane north
minneapolis MN 55442
MDR Report Key7389942
Report Number3012307300-2018-00899
Device Sequence Number1417565
Product Code FOZ
UDI-Device Identifier10351688077851
UDI-Public10351688077851
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K113700
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor,foreign,health pr
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/30/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/22/2019
Device Catalogue Number326800
Device Lot Number3417549
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/13/2018
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/08/2018
Supplement Date Received by Manufacturer04/18/2018
Initial Report FDA Received Date04/02/2018
Supplement Report FDA Received Date05/30/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-