| Model Number |
E SERIES |
| Medical Device Problem Code |
No Display/Image (1183)
|
| Health Effect - Clinical Code |
No Information (3190)
|
| Date of Event |
03/07/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation has received the product and will be providing a supplemental report when our investigation is completed.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that while attempting to treat a (b)(6) year old female patient, the device's display was distorted and clinical information could not be seen.Complainant did not indicate that there was any adverse effect to the patient due to the reported malfunction.Complainant indicated that during subsequent functional testing, the malfunction was not duplicated.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned to zoll medical corporation and the customer's report was not replicated or confirmed.The device was put through extensive testing without duplicating the report.The video cable was replaced as a precaution.The device was recertified and returned to the customer.No trend is associated with reports of this type.
|
| |
|