Model Number 48335310
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 03/05/2018
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the locking ring on the screw disengaged upon insertion.The affected screw was replaced with a new one, with a 5 minute surgical delay.There are no reports of adverse health consequences to the patient.
 
Additional Manufacturer Narrative
Results: visual, dimensional and functional analysis could not be performed as the device was not returned.No lot # was provided, so a manufacturing record review could not be performed.Conclusion: the definitive root cause of the event cannot be determined from the given information and without device evaluation.Potential root causes of the locking ring detachment may include: freehand use during screw insertion, inserter loosening during surgery, inserting not fully threading, user unfamiliar with system, interaction between cage/screw/inserter, outer diameter of clip is too large, general use error.
 
Event or Problem Description
It was reported that the locking ring on the screw disengaged upon insertion.The affected screw was replaced with a new one, with a 5 minute surgical delay.There are no reports of adverse health consequences to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANCHOR C DIAM.3.5MM SELF DRILLING 10MM
Common Device Name
INTERVERTEBRAL FUSION DEVICE WITH INTEGRATED FIXATION, CERVICAL
Manufacturer (Section D)
STRYKER SPINE-US
2 pearl court
-
allendale NJ 07401
MDR Report Key7390578
Report Number3005525032-2018-00017
Device Sequence Number13761506
Product Code OVE
UDI-Device Identifier04546540679833
UDI-Public(01)04546540679833
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K102606
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/25/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number48335310
Device Catalogue Number48335310
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 03/05/2018
Supplement Date Received by Manufacturer05/29/2018
Initial Report FDA Received Date04/02/2018
Supplement Report FDA Received Date06/25/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age62 YR
-
-