• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER SPINE-US XIA ROD TO ROD CLAMP - AXIAL; THORACOLUMBOSACRAL PEDICLE SCREW SYSTEM22

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER SPINE-US XIA ROD TO ROD CLAMP - AXIAL; THORACOLUMBOSACRAL PEDICLE SCREW SYSTEM22 Back to Search Results
Model Number 03805002
Medical Device Problem Code Migration (4003)
Health Effect - Clinical Codes Pain (1994); Injury (2348)
Date of Event 03/08/2018
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a patient presented increasing back pain at 1 year out meeting.X-rays showed that both rods were dislodged from the axial connectors at the l1 level.Revision surgery has been scheduled for the patient.
 
Additional Manufacturer Narrative
Lot #: 14a084.The reported event was confirmed via x-rays and visual inspection.It was reported that the rods disengaged from the connectors one year after surgery.Manufacturing records were reviewed for the corresponding lot and no relevant issues were identified.Per device inspection, the opening was heavily deformed indicating that the rod was pushed down onto the bottom of the opening, deforming the connector.The set screws were examined.It was found that the set screws had heavy rod indentations on the bottom, indicating that excess force was applied to the set screw.The bottom threading of the set screws were also found to be deformed, most likely from excessive force being applied and pushing the connector against the rod.Inspection of the device indicates that excessive force was applied to the set screw, which pushed against the rod, which in turn pushed against the connector opening, deforming both the set screw and the connector itself.This caused instability in the connector as the set screw and connector are not properly securing the rod in place due to the deformation.Over the course of the year it was implanted, the rod worked it's way out of the deformed connector, most likely due to the patient's movement and eventually disengaged completely.The most likely root cause was excessive force applied during set screw tightening.
 
Event or Problem Description
It was reported that a patient presented increasing back pain at 1 year out meeting.X-rays showed that both rods were dislodged from the axial connectors at the l1 level.Revision surgery has been scheduled for the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XIA ROD TO ROD CLAMP - AXIAL
Common Device Name
THORACOLUMBOSACRAL PEDICLE SCREW SYSTEM22
Manufacturer (Section D)
STRYKER SPINE-US
2 pearl court
-
allendale NJ 07401
MDR Report Key7390661
Report Number3005525032-2018-00022
Device Sequence Number16456976
Product Code NKB
UDI-Device Identifier04546540147998
UDI-Public(01)04546540147998
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K060361
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/07/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Model Number03805002
Device Catalogue Number03805002
Device Lot Number14A084
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/29/2018
Initial Date Received by Manufacturer 03/09/2018
Supplement Date Received by Manufacturer07/17/2018
Initial Report FDA Received Date04/02/2018
Supplement Report FDA Received Date08/07/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age75 YR
Patient Weight102
-
-