PHYSIO-CONTROL, INC. - 3015876 LIFEPAK® 1000 DEFIBRILLATOR; AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE) Back to Search Results
Model Number 1000
Medical Device Problem Code Inappropriate/Inadequate Shock/Stimulation (1574)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 02/06/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Physio-control evaluated the customer's device and was able to verify the reported issue.Physio determined that the cause of the reported issue was due to the analog pcb assembly.A therapy harness ferrite bead was placed over integrated circuit, designator u1, which broke the solder connections at legs 1 and 10 and caused excessive voltage on discharge which burned resistor, designator r212.
 
Event or Problem Description
The customer contacted physio-control to report that their device displayed the service indicator and was alarming.There was no patient use associated with the reported event.Upon evaluating the customer's device, physio observed a failure that resulted in a partial loss of defibrillator output energy due to a loss of the negative portion of the biphasic output waveform, resulting in a monophasic shock.As a result, the wrong defibrillation therapy would be delivered, if it were necessary.
 
Event or Problem Description
The customer contacted physio-control to report that their device displayed the service indicator and was alarming.There was no patient use associated with the reported event.Upon evaluating the customer's device, physio observed a failure that resulted in a partial loss of defibrillator output energy due to a loss of the negative portion of the biphasic output waveform, resulting in a monophasic shock.As a result, the wrong defibrillation therapy would be delivered, if it were necessary.
 
Additional Manufacturer Narrative
Section d4 of the initial medwatch report indicates: device identifier (di) number ¿ blank.Section d4 of the initial medwatch report should indicate: device identifier (di) number ¿ (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEPAK® 1000 DEFIBRILLATOR
Common Device Name
AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)
Manufacturer (Section D)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
MDR Report Key7390716
Report Number0003015876-2018-00529
Device Sequence Number785804
Product Code MKJ
UDI-Device Identifier00885074698878
UDI-Public00885074698878
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K122600
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source user facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 02/05/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1000
Device Catalogue Number99425-000023
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/13/2018
Initial Date Received by Manufacturer 03/09/2018
Supplement Date Received by Manufacturer01/16/2020
Initial Report FDA Received Date04/02/2018
Supplement Report FDA Received Date02/05/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-