| Model Number |
1000 |
| Medical Device Problem Code |
Inappropriate/Inadequate Shock/Stimulation (1574)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
02/06/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Physio-control evaluated the customer's device and was able to verify the reported issue.Physio determined that the cause of the reported issue was due to the analog pcb assembly.A therapy harness ferrite bead was placed over integrated circuit, designator u1, which broke the solder connections at legs 1 and 10 and caused excessive voltage on discharge which burned resistor, designator r212.
|
| |
|
Event or Problem Description
|
|
The customer contacted physio-control to report that their device displayed the service indicator and was alarming.There was no patient use associated with the reported event.Upon evaluating the customer's device, physio observed a failure that resulted in a partial loss of defibrillator output energy due to a loss of the negative portion of the biphasic output waveform, resulting in a monophasic shock.As a result, the wrong defibrillation therapy would be delivered, if it were necessary.
|
| |
|
Event or Problem Description
|
|
The customer contacted physio-control to report that their device displayed the service indicator and was alarming.There was no patient use associated with the reported event.Upon evaluating the customer's device, physio observed a failure that resulted in a partial loss of defibrillator output energy due to a loss of the negative portion of the biphasic output waveform, resulting in a monophasic shock.As a result, the wrong defibrillation therapy would be delivered, if it were necessary.
|
| |
|
Additional Manufacturer Narrative
|
|
Section d4 of the initial medwatch report indicates: device identifier (di) number ¿ blank.Section d4 of the initial medwatch report should indicate: device identifier (di) number ¿ (b)(4).
|
| |
|