| Model Number |
MMT-523NAL |
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
03/12/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
The customer reported via phone call that the insulin pump had partial display and the screen is not legible or it is very faint.The customer¿s blood glucose level was 212 mg/dl at the time of the incident.The customer was assisted with troubleshooting.The customer was advised to discontinue use of the insulin pump and to revert to the back-up plan.The insulin pump will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Device received with missing segments or partial display due to cracked lcd zebra connector.Unable to perform any tests due to missing segments or partial display.
|
| |
|