• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK INC FLEXOR RTPS GUIDING SHEATH; DYB INTRODUCER, CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK INC FLEXOR RTPS GUIDING SHEATH; DYB INTRODUCER, CATHETER Back to Search Results
Model Number N/A
Medical Device Problem Code Split (2537)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 03/12/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).510k status: preamendment.This report includes information known at this time.A follow up report will be submitted should additional relevant information become available.
 
Event or Problem Description
The customer reported that, when the package of the flexor rtps guiding sheath was opened for a transjugular liver biopsy, it was discovered that the dilator that was included with the introducer had an irregular tip; there appeared to be a split in the tip of the device.The operator attempted to introduce it through the skin of the patient, but subsequently ran into unspecified difficulties.The operator replaced the product with a new device, and no further issues were reported.No anatomical abnormalities were noted at the access site.The customer confirmed that no patient adverse events resulted from the product problem, and no additional procedures were necessitated.The complaint device is reportedly unavailable for return and evaluation.
 
Additional Manufacturer Narrative
Investigation - evaluation: a review of the complaint history, device history record, documentation, drawing, manufacturing instructions, quality control, and specifications of the device was conducted during the investigation.The complaint device was not returned; therefore, no physical examinations could be performed.However, a document based investigation evaluation was performed.There is no evidence to suggest the product was not made to specifications.The appropriate manufacturing documents from the device master record (dmr) pertaining to the reported failure mode were reviewed, revealing no notable gaps in the manufacturing process.Documentation indicates to confirm the distal tip of the dilator is rounded, smooth, free of excessive nicks and debris, and to reject the product if any splits are noted.A review of the device history record shows no nonconforming events which could contribute to this failure mode.It should be noted there were no other reported complaints for this lot number.There is no evidence confirming this failure was manufacturing related or shipping/handling related.Based on the information provided, no product returned and the results of our investigation, a definitive root cause could not be determined.We will continue to monitor for similar complaints and have notified the proper personnel about this event.Per the quality engineering risk assessment no further action is required.
 
Event or Problem Description
No additional information regarding the patient and/or event has been received since the previous medwatch report was sent.
 
Additional Manufacturer Narrative
Blank fields on this form indicate the information is unknown, unavailable, or unchanged.Correction: b3, d1, d4, g5, h6, h8, h10.G5-pma/510(k) = k142829.Event summary it was reported that, when the package of the flexor rtps guiding sheath was opened for a transjugular liver biopsy, the tip of the dilator was observed split.The user unsuccessfully attempted to introduce the device before subsequently replacing it.No anatomical abnormalities were noted at the access site.The customer confirmed that no patient adverse events resulted from the product problem, and no additional procedures were necessary.Investigation - evaluation a review of the complaint history, device history record, documentation, drawing, manufacturing instructions, quality control, and specifications of the device was conducted during the investigation.The complaint device was not returned; therefore, no physical examinations could be performed.However, a document-based investigation evaluation was performed.The appropriate manufacturing documents from the device master record (dmr) pertaining to the reported failure mode were reviewed, revealing no notable gaps in the manufacturing process.Documentation indicates to confirm the distal tip of the dilator is rounded, smooth, free of excessive nicks and debris, and to reject the product if any splits are noted.A review of the device history record shows no nonconforming events which could contribute to this failure mode.It should be noted there were no other reported complaints for this lot number.The complaint device is shipped with instructions for use (ifu), t_intro_rev2, which states, "upon removal from package, inspect the product to ensure no damage has occurred.¿ based on the information provided, no product returned and the results of our investigation, a definitive root cause for the device failure could not be determined.Possible causes for the device failure include incorrect tipping during manufacturing or damage incurred during shipping and handling.As no gaps in the manufacturing or quality control process were identified, however, there is objective evidence to suggest that the device was manufactured to specifications.We will continue to monitor for similar complaints and have notified the proper personnel about this event.Per the quality engineering risk assessment, no further action is required.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLEXOR RTPS GUIDING SHEATH
Common Device Name
DYB INTRODUCER, CATHETER
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
MDR Report Key7390814
Report Number1820334-2018-00858
Device Sequence Number1677951
Product Code DYB
UDI-Device Identifier00827002130814
UDI-Public(01)00827002130814(17)210126(10)8538789
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,user f
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 03/10/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue NumberKCFW-10.0-35-RB-RTPS-100
Device Lot Number8538789
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 03/12/2018
Supplement Date Received by Manufacturer05/04/2018
03/06/2020
Initial Report FDA Received Date04/02/2018
Supplement Report FDA Received Date05/31/2018
03/10/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-