| Catalog Number |
AB35W06030135 |
| Medical Device Problem Code |
Burst Container or Vessel (1074)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
03/05/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that the physician used an evercross device during treatment of a moderately calcified lesion in the mid left superficial femoral artery (sfa).The lesion was reported to exhibit 90% stenosis and the vessel was reported to exhibit no tortuosity.The device was prepped as per ifu without issue.No embolic protection was used and the device did not pass through a previously deployed stent.No resistance was encountered during advancement of the device.A non-mdt inflation device was used.It is reported that a balloon burst occurred during balloon inflation at a pressure of 8atms on the second inflation.There were no balloon fragments.A 6x40 everflex stent was then deployed to cover the lesion.No patient injury was reported.
|
| |
|