COVIDIEN MFG DC BOULDER LIGASURE; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Model Number LS3112
Medical Device Problem Code Failure to Fire (2610)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 11/20/2017
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Evaluation summary: one device was received for evaluation.This device had been used in the treatment or diagnosis of a patient.A review of the lot number reported indicates that the product was within the assigned expiration date at the time of the reported incident.The returned product did not meet specification as received.Visual inspection found one of the snaps to have broken off.The broken piece was not returned.The reported condition was not confirmed.The device's electrical resistance was found to met specification.The disposable was assembled onto a set of forceps and plugged into a forcetriad generator.The generator recognized the device and activated with satisfactory results.All seal cycles were completed satisfactorily and end tones were heard indicating completed activation cycles.The investigation found the device to activate normally.An additional failure was found during the investigation; one of the snaps was found to be broken and missing.The additional findings did not contribute to the reported condition.The investigation found a broken snap on the electrode jaw.Snap damage can be caused by the user improperly misaligning the snaps into the mating holes during assembly.This damage can also be caused by multiple assembly attempts.The investigation identified the root cause of the reported event to be attributed to user error.Refer to the product instructions for use for additional information on the proper method of assembling the electrodes onto the reusable instrument.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
The customer reported that during the procedure at the time of initial sealing, the device was not activated by pressing the activation button.A new device was used to continue.There was no patient harm.Manufacturer's initial evaluation found one of the snaps to have broken off.The broken piece was not returned.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIGASURE
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
COVIDIEN MFG DC BOULDER
5920 longbow dr
boulder CO 80301 3299
MDR Report Key7391939
Report Number1717344-2018-00453
Device Sequence Number1160584
Product Code GEI
UDI-Device Identifier20884524002511
UDI-Public20884524002511
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K043273
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,user facility
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 04/03/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/30/2022
Device Model NumberLS3112
Device Catalogue NumberLS3112
Device Lot Number71510250X
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/21/2017
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/12/2018
Initial Report FDA Received Date04/03/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-