The marathon microcatheter was returned for analysis.Upon visual inspection, no issues were found with the marathon micro catheter hub.No kinks or bends were found with the body.The marathon micro catheter distal tip was found to be shaped which is consistent with the reported information.A hole in the tubing material was found at the proximal edge of the marker band.The hole appears to have been caused by melting.The marathon micro catheter was flushed, water exited from the hole and distal tip.Based on the device analysis and reported information, the report of catheter damage (hole) was confirmed.Although the root cause could not be determined, it is likely the distal tip became damaged during the shaping process.The lot history record review of the reported lot number showed no discrepancies that would have contributed to the reported experience; manufacturing has been ruled out as a potential cause.All products are 100% inspected for damages and irregularities during manufacture.The marathon ifu (instructions for use) provides the following guidance: ¿remove shaping mandrel from card and insert into distal tip of the catheter.Carefully bend catheter tip and shaping mandrel to desired shape.A slight over exaggeration of the shape may be required to accommodate for catheter relaxation.Shape the catheter by holding the shaped portion approximately 1 inch (2.5 cm not less than 1 cm) from the steam source for about 20 seconds (do not exceed 30 seconds).Allow catheter tip to cool in air or saline prior to removing mandrel.Remove mandrel from catheter and discard.Multiple shaping is not recommended.Inspect the tip for any damage that may have resulted from steam shaping the catheter.If any damage is found, do not use the catheter.¿ if information is provided in the future, a supplemental report will be issued.
|