Catalog Number RBY2C0312
Medical Device Problem Codes Failure to Advance (2524); Physical Resistance (2578)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 03/05/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Results: the pet lock was intact on the proximal end of the pusher assembly.The pusher assembly was kinked in multiple locations throughout its length.The embolization coil was intact with the pusher assembly and compressed throughout its length, with the stretch resistant wire (sr wire) wrapped around the distal detachment tip (ddt).Conclusions: evaluation of the returned device revealed the embolization coil was compressed.This type of damage typically occurs due to forceful advancement of the ruby coil against resistance.The initial cause of resistance could not be determined, but the compression of the embolization coil prevented the ruby coil from advancing out of its introducer sheath during functional testing.Further evaluation revealed the pusher assembly was kinked in multiple locations.This damage was likely incidental and may have occurred during packaging for return to penumbra.The non-penumbra microcatheter referenced in the complaint was not returned for evaluation.Penumbra coils are inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
 
Event or Problem Description
The patient was undergoing a coil embolization procedure in the gastroduodenal artery (gda) using ruby coils.During the procedure, the physician felt resistance while advancing a ruby coil through a non-penumbra microcatheter, and therefore the ruby coil was removed.The procedure was completed using two additional ruby coils and the same microcatheter.There was no report of an adverse effect to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RUBY COIL
Common Device Name
HCG, KRD
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key7392263
Report Number3005168196-2018-00672
Device Sequence Number1092771
Product Code HCG
UDI-Device Identifier00814548012827
UDI-Public00814548012827
Combination Product (Y/N)N
Initial Reporter StateID
Initial Reporter CountryUS
PMA/510(K) Number
K120330
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 03/06/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Expiration Date11/16/2023
Device Catalogue NumberRBY2C0312
Device Lot NumberF66540
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/13/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/06/2018
Initial Report FDA Received Date04/03/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age74 YR
-
-