| Model Number |
TRUMETRXGO |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Code |
Fatigue (1849)
|
| Date of Event |
03/09/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Product not yet returned for evaluation.Most likely underlying root cause: mlc-58-user had an inaccurate reference: self: the person is using themselves as the reference or how they feel at the time they run the blood test.Test strip udi# (b)(4).Manufacturer contacted customer on (b)(6) in a follow-up call in order to ensure the customer's condition since the initial call; customer condition improved.No symptoms or medical attention reported since the original call.Customer satisfied with the replacement product.
|
| |
|
Event or Problem Description
|
|
Consumer reported complaint for high blood glucose results.The customer is concerned with results obtained of 308 mg/dl.The expected fasting blood glucose test result range is 140 to 150 mg/dl.The customer and did report symptom of feeling tired.Medical attention is not reported as a result of the actual blood glucose results and reported symptoms.The product is stored by customer according to specification in the bedroom.During the call on (b)(6) 2018, a back to back blood test was refused by the customer.The test strip lot manufacturer's expiration date is 03/28/2018 and open vial date is (b)(6) 2017.The meter memory was reviewed for previous test result history: (b)(6).Customer called in states the meter is reading high.Customer states the meter read 308 mg/dl fasting.Customer states his normal fasting range is 140-150 mg/dl fasting.Customer denies the need for medical attention at the time of call and denies signs and symptoms of hyperglycemia at the time of call.
|
| |
|
Search Alerts/Recalls
|