Model Number 27966
Medical Device Problem Code Power Problem (3010)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was returned to an authorized resmed third party service center and an evaluation confirmed the complaint.The power supply unit was replaced to address this issue.The device was serviced and fully tested before it was returned to the customer.Resmed reference #: (b)(4).
 
Event or Problem Description
It was reported to resmed that an astral power supply unit (psu) was inoperable.There was no patient harm or serious injury reported as a result of this incident.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ASTRAL PSU SLIM LINE
Manufacturer (Section D)
RESMED LTD
1 elizabeth macarthur drive.
bella vista
sydney, nsw 2153
AU  2153
MDR Report Key7392887
Report Number3004604967-2018-00545
Device Sequence Number1677889
Product Code CBK
UDI-Device Identifier00619498279661
UDI-Public(01)00619498279661(10)1253994
Combination Product (Y/N)N
Initial Reporter StateAL
Initial Reporter CountryUS
PMA/510(K) Number
K152068
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor
Initial Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial
Report Date (Section B) 04/03/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Medical Equipment Company Technician/Representative
Device Model Number27966
Device Catalogue Number27966
Was Device Available for Evaluation? No
Distributor Facility Aware Date03/14/2018
Device Age7 MO
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/14/2018
Initial Report FDA Received Date04/03/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-