• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC - SPENCER ADVANIX¿ PANCREATIC STENT; CATHETER, BILIARY, DIAGNOSTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC - SPENCER ADVANIX¿ PANCREATIC STENT; CATHETER, BILIARY, DIAGNOSTIC Back to Search Results
Model Number M00536760
Medical Device Problem Code Kinked (1339)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 03/16/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).According to the complainant, the suspect device has been disposed and is not available for return.If any further relevant information is received, a supplemental mdr will be filed.
 
Event or Problem Description
It was reported to boston scientific corporation that an advanix pancreatic stent pigtail was used during an endoscopic retrograde cholangiopancreatography (ercp) and stent placement procedure performed in the pancreatic duct on (b)(6) 2018.According to the complainant, during the procedure, the physician loaded the pusher on and attempted to deploy the advanix 3f plastic stent, however, it was noticed that the stent coiled up into the pusher.The procedure was completed with a 4f advanix stent.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be fine.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVANIX¿ PANCREATIC STENT
Common Device Name
CATHETER, BILIARY, DIAGNOSTIC
Manufacturer (Section D)
BOSTON SCIENTIFIC - SPENCER
780 brookside drive
spencer IN 47460
MDR Report Key7392957
Report Number3005099803-2018-00981
Device Sequence Number1160578
Product Code FGE
UDI-Device Identifier08714729856962
UDI-Public08714729856962
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K133700
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 03/16/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/19/2019
Device Model NumberM00536760
Device Catalogue Number3676
Device Lot Number21157182
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/16/2018
Initial Report FDA Received Date04/03/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-