MAUDE Adverse Event Report: DATEX-OHMEDA, INC. CARESCAPE R860; VENTILATOR, CONTINUOUS, FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATEX-OHMEDA, INC. CARESCAPE R860; VENTILATOR, CONTINUOUS, FACILITY USE Back to Search Results
Device Problems Defective Alarm (1014); Appropriate Term/Code Not Available (3191)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 03/06/2018
Event Type  malfunction  
Manufacturer Narrative
A ge healthcare service representative performed a checkout of the system and confirmed the reported issue.The ventilator control board and the power management board were replaced to resolve the reported issue.No report of patient involvement.(b)(4).The initial reporter is located outside the u.S., and therefore this information is not provided due to country privacy laws.
 
Event Description
The hospital reported that when the device was connected to a patient, it shut down by itself without issuing any alarm and mechanical ventilation stopped.There was no report of patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARESCAPE R860
Type of Device
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
DATEX-OHMEDA, INC.
3030 ohmeda dr,
madison, WI
Manufacturer Contact
john szalinski
3000 n grandview blvd.
waukesha, WI 
MDR Report Key7393107
MDR Text Key104253069
Report Number2112667-2018-00677
Device Sequence Number1
Product Code CBK
Combination Product (y/n)N
PMA/PMN Number
K151570
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,user facility
Type of Report Initial
Report Date 04/03/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/03/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Other Device ID NumberSEE BLOCK H10
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received03/06/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/18/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-