• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE - ENGLEWOOD 3000 ML TPN BAG; SET, I.V. FLUID TRANSFER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE - ENGLEWOOD 3000 ML TPN BAG; SET, I.V. FLUID TRANSFER Back to Search Results
Catalog Number H938741
Device Problems Fluid/Blood Leak (1250); Material Rupture (1546)
Patient Problem No Patient Involvement (2645)
Event Date 03/08/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(6).Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that a 3000ml exactamix eva tpn bag had ruptured and leaked from right under the hanger hole.This event occurred while the tubing was being attached to the bag during set up.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
The user facility submitted voluntary event report mw5075944 for this event.The actual device and a companion sample were received for evaluation.Visual inspection of the actual device revealed a large tear at the top middle section of the bag.The tear was approximately 4 inches in length.The reported condition was verified for the actual sample.The cause of the condition was not determined.To address this issue, the supplier of the component has been notified of the condition.Visual inspection of the companion sample did not identify any abnormalities that could have contributed to the reported condition.The reported condition was not verified for the companion sample.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
3000 ML TPN BAG
Type of Device
SET, I.V. FLUID TRANSFER
Manufacturer (Section D)
BAXTER HEALTHCARE - ENGLEWOOD
englewood CO
Manufacturer (Section G)
BAXTER HEALTHCARE - ENGLEWOOD
14445 grasslands dr
englewood CO 80112
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key7393133
MDR Text Key104424205
Report Number1416980-2018-01819
Device Sequence Number1
Product Code LHI
UDI-Device Identifier00085412477251
UDI-Public(01)00085412477251
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K900585
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 05/03/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/03/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date05/31/2020
Device Catalogue NumberH938741
Device Lot Number60066290
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/03/2018
Is the Reporter a Health Professional? No
Date Manufacturer Received04/20/2018
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-