Blank fields on this form indicate the information is unknown, unavailable, or unchanged.Investigation evaluation.A review of the complaint history, device history record, documentation, drawing, quality control, specifications, and a visual inspection of the returned devices were conducted during the investigation.The visual inspection of the returned devices confirmed that two of the returned samples contained white/yellow substances on their respective coatings.Additionally, a document based investigation evaluation was performed.A review of the device history record showed the lot was released meeting all finished goods release criteria.Due to the individual nature of the product manufacturing and 100% verification of various defects including inspection of the device surface in final qc prior to shipping, there is no indication that there is nonconforming product in the field.It should be noted there were no other reported complaints for this lot number.Based on the information provided, the examination of the returned product, and the results of our investigation, a definitive root cause could not be determined.We will continue to monitor for similar complaints.Per the quality engineering risk assessment, no further action is required.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|