BAXTER HEALTHCARE - HECHINGEN BADEN DIALYZER REVACLEAR 400; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM Back to Search Results
Catalog Number 114746
Medical Device Problem Codes Disconnection (1171); Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 03/08/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(6).Should additional relevant information become available, a supplemental report will be submitted.
 
Event or Problem Description
Reportedly, a blood leak was detected 3 hours after the treatment began, at the connection of the revaclear 400 dialyzer to the arterial dialyzer line.The presence of coagulated blood was noted at the level of the inlet and outlet connectors of the dialyzer.According to the reporter, the arterial expansion chamber was under high pressure, a clot in the line was noticed, which in turn led to an overpressure on the venous line.The session was discontinued.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
 
Additional Manufacturer Narrative
A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.The device was not received for evaluation; therefore, a device analysis could not be completed.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REVACLEAR 400
Common Device Name
DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
BAXTER HEALTHCARE - HECHINGEN BADEN DIALYZER
hechingen
MDR Report Key7393468
Report Number9611369-2018-00032
Device Sequence Number13739158
Product Code KDI
UDI-Device Identifier07332414114923
UDI-Public(01)07332414114923
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K130039
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Pharmacist
Type of Report Initial,Followup
Report Date (Section B) 05/18/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/30/2020
Device Catalogue Number114746
Device Lot Number7-7087-H-01
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 03/08/2018
Supplement Date Received by Manufacturer05/18/2018
Initial Report FDA Received Date04/03/2018
Supplement Report FDA Received Date05/18/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
AK 200 ULTRA S DIALYSIS MACHINE; NIPRO LINES
-
-