WRIGHTS LANE SYNTHES USA PRODUCTS LLC AIM-ARM RADIOLUC F/EXPERT ALFN; ROD,FIXATION,INTRAMEDULLARY
|
Back to Search Results |
|
| Model Number |
03.010.483 |
| Medical Device Problem Code |
Failure to Align (2522)
|
| Health Effect - Clinical Code |
No Code Available (3191)
|
| Date of Event |
03/07/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Patient identifier, date of birth, age, gender and weight were not available for reporting.Device is an instrument and is not implanted / explanted.Reporter title is orthopedics coordinator.The device was received and the product evaluation is in progress.No conclusion can be drawn.Patient code (b)(4) used to capture additional medical/surgical intervention required due to change in surgical plan.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that the screws would not go in the aiming arm when the surgeon was attempting to insert the proximal locking screws in an adolescent femoral nail during an intramedullary nail of a femur on (b)(6) 2018.The sales consultant further explained that the aiming arm would accept the drill bit, but not the screws.The insertion handle and/or the aiming arm kept the protection sleeve from lining up properly, which eventually led to the transverse interlock not working properly and they had to switch the hole and use the antegrade interlock to lock the screw and complete the procedure.The procedure was successfully completed with no surgical delay.The patient was fine at the end of the procedure.Concomitant device (s) reported:12.0mm/8.0mm protection sleeve (part # 03.010.063, unknown lot #, qty # 1).This report is for one (1) radiolucent recon aiming arm.This is report 2 of 3 for (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Udi: (b)(4).A device history record review was performed for the subject device: part number: 03.010.483, synthes lot number: t987427, release to warehouse date: 12-mar-2013, manufacture site: tuttlingen.Part expiration date: n/a.List of nonconformance¿s: 112690.An non-conformance (ncr) was started because design was changed (the color mark was changed from green to blue).The parts were reworked and were checked 100% and found to be good.The ncs have no impact on the complaint issue.Further review of the device history records showed that there were no issues at the time of manufacturing of this device and it's sub components that would contribute to the complaint condition.Review of the device history record of tuttlingen showed that there were no issues at the time of manufacturing of this device that would contribute to the issue outlined in this complaint.A product investigation was performed for the subject device: it was reported that the screws would not go in the aiming arm when the surgeon was attempting to insert the proximal locking screws in an adolescent femoral nail during an intramedullary nail of a femur on (b)(6) 2018.The sales consultant further explained that the aiming arm would accept the drill bit, but not the screws.The insertion handle and/or the aiming arm kept the protection sleeve from lining up properly, which eventually led to the transverse interlock not working properly and they had to switch the hole and use the antegrade interlock to lock the screw and complete the procedure.The procedure was successfully completed with no surgical delay.The patient was fine at the end of the procedure.The reported complaint condition of device interaction could not be confirmed or replicated at customer quality (cq) because only the aiming arm and insertion handle were returned.Post manufacturing damage in the form of a burr was observed on the inferior side of the insertion handle hole.Therefore, this complaint is confirmed for post manufacturing damage.The returned aiming arm was manufactured in march 2013 and is over 5 years old.Review of the device history record(s) showed that there were no issues during the manufacture of the product that would contribute to this complaint condition.The material and relevant material properties were determined to be conforming at the time of manufacture based on review of the dhr.The following drawing was reviewed during investigation.Aiming arm ¿ the design history was not found to impact the complaint condition.During the investigation no product design issues or discrepancies were observed that may have contributed to the complaint condition.The returned parts were determined to be suitable for the intended use when employed.The radiolucent aiming arm for expert alfn is part of the titanium cannulated adolescent lateral femoral nail ¿ ex insertion instrument set (01.116.044).When fully assembled as an aiming arm construct, the radiolucent aiming is a multi-use device that is used to guide recon and standard screws into the adolescent lateral entry femoral nail.A dimensional inspection was not performed at cq because the returned 5+ year old reusable device assembled as intended to the returned insertion handle and no damage was identified.The recon femoral nailing implants - core dossier design & clinical risk management (dcrm) was reviewed and the dcrm adequately addresses this complaint.No design issues or manufacturing discrepancies were identified during this investigation.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|