| Model Number |
012001 |
| Medical Device Problem Codes |
Failure to Cut (2587); Noise, Audible (3273)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
03/02/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during laparoscopic living donor nephrectomy, metallic noise was heard while in use, and the device did not cut tissue anymore.There was no patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation summary: post market vigilance (pmv) led an evaluation of one device.The instrument registered a mechanical fault.The instrument was disass embled, a crack was noticed in the proximal portion of the probe shaft.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.Replication of the cracked probe shaft may occur when the probe shaft makes contact with the outer tube while the system was energized.The information for use for this product states, avoid using the device as a lever, either while dissecting or while activating the instrument.Added stress to the tip may cause the probe to touch in the inner tube portion of the instrument, resulting in mechanical failure and/or rendering the instrument inoperable.The root cause of the observed damage was misuse of the product which would have caused or contributed to the reported incident.A review of the current complaint data reveals no trend for a device related failure for this condition.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during laparoscopic living donor nephrectomy, metallic noise was heard while in use, and the device did not cut tissue anymore.New device was opened to complete the case.There was no patient injury.
|
| |
|