| Catalog Number |
RSINT30030UX |
| Medical Device Problem Code |
Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
03/08/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that the physician intended to use resolute integrity rx drug eluting stent to treat a moderately tortuous and calcified lesion located in the mid right coronary artery, exhibiting 90% stenosis.There were no abnormalities reported in relation to anatomy.There was no damage noted to packaging.There were no issues noted when removing the device from the hoop/tray.The device was inspected with no issues identified.Negative prep was not performed.The lesion was not pre-dilated.The device did not pass through a previously-deployed stent.Resistance was encountered when advancing the device.Excessive force was used during delivery.It was reported that the stent deformation occurred during positioning/advancement.It was reported that the stent caught on the guideliner during advancement.Patient status post-procedure is alive with no injury.The procedure was completed with a stent of the same size.
|
| |
|