MEDTRONIC IRELAND RESOLUTE INTEGRITY RX; STENT, CORONARY, DRUG-ELUTING Back to Search Results
Catalog Number RSINT30030UX
Medical Device Problem Code Material Deformation (2976)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 03/08/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that the physician intended to use resolute integrity rx drug eluting stent to treat a moderately tortuous and calcified lesion located in the mid right coronary artery, exhibiting 90% stenosis.There were no abnormalities reported in relation to anatomy.There was no damage noted to packaging.There were no issues noted when removing the device from the hoop/tray.The device was inspected with no issues identified.Negative prep was not performed.The lesion was not pre-dilated.The device did not pass through a previously-deployed stent.Resistance was encountered when advancing the device.Excessive force was used during delivery.It was reported that the stent deformation occurred during positioning/advancement.It was reported that the stent caught on the guideliner during advancement.Patient status post-procedure is alive with no injury.The procedure was completed with a stent of the same size.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESOLUTE INTEGRITY RX
Common Device Name
STENT, CORONARY, DRUG-ELUTING
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway
MDR Report Key7395028
Report Number9612164-2018-00628
Device Sequence Number128293
Product Code NIQ
UDI-Device Identifier00643169037182
UDI-Public00643169037182
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
P110013
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 04/04/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/19/2019
Device Catalogue NumberRSINT30030UX
Device Lot Number0008591119
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/08/2018
Initial Report FDA Received Date04/04/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age70 YR
Patient Weight96
-
-