Model Number ETLW1610C124E
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Codes Pain (1994); Injury (2348)
Date of Event 03/08/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
An endurant ii stent graft system was implanted in a patient for the endovascular treatment of a 50mm diameter abdominal aortic aneurysm.It was reported that, approximately 4 years post index procedure the patient presented with abdominal tenderness.On ct exam there appeared to be a type ia endoleak.The next day, after endoanchors deployment for the type ia endoleak, the angiogram revealed a jetting endoleak from the wall of the main body of the bifurcated device.An aortic cuff was deployed and sealed the hole but on the next angiogram another hole was observed in the contralateral limb.The physician relined the limb with a 16x16x82 endurant limb and the final angiogram showed the endoleak had resolved.As per the physician, the cause of the event cannot be determined.No additional clinical sequelae were reported and the patient is fine.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDURANT II ILIAC STENT GRAFT
Common Device Name
SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway
MDR Report Key7395029
Report Number2953200-2018-00472
Device Sequence Number1273804
Product Code MIH
UDI-Device Identifier00613994991096
UDI-Public00613994991096
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P100021
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2014
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 04/04/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/10/2015
Device Model NumberETLW1610C124E
Device Catalogue NumberETLW1610C124E
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/09/2018
Initial Report FDA Received Date04/04/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-