Other relevant device(s) are: product id: vamf3636c200tu, serial/lot #: (b)(4), ubd: (b)(6) 2016, udi#: (b)(4) ; product id: vamf3838c100tu, serial/lot #: (b)(4), ubd: (b)(6) 2016, udi#: (b)(4).If information is provided in the future, a supplemental report will be issued.
|
|
A valiant stent graft system was implanted in a patient for the endovascular treatment of an 80mm thoracic aortic aneurysm.It was reported that the patient presented emergently approximately 4 years post index procedure.A type iii endoleak was discovered on a ct scan.The patient was visiting from michigan, where the original stent graft had been implanted, when he started experiencing the back pain.The physician did a brachial and femoral artery exposure and a vamf3838c200te was implanted inside of the existing grafts to cover the type iii endoleak.The grafts were repaired.No further clinical sequelae were reported and the patient is fine.
|