MAUDE Adverse Event Report: MEDTRONIC IRELAND RESOLUTE ONYX RX; STENT, CORONARY, DRUG-ELUTING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC IRELAND RESOLUTE ONYX RX; STENT, CORONARY, DRUG-ELUTING Back to Search Results
Catalog Number RONYX40022UX
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code Patient Problem/Medical Problem (2688)
Date of Event 03/09/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that the physician intended to use a resolute onyx drug eluting stent to treat a severely calcified and mildly tortuous lesion, exhibiting 98% lesion stenosis between the mid to proximal saphenous vein graft.There was no damage noted to packaging, i.E.Shelf carton, hoop/tray.There were no issues noted when removing the device from the hoop/tray.The device was inspected with no issue identified.Negative prep was also performed with no issues.The lesion was not pre-dilated.The device did pass through a previously-deployed stent.Resistance was encountered when advancing the device.Excessive force was not used during delivery.It was reported that stent dislodgement occurred during removal following failed delivery.The dislodged stent was removed using a snare.Patient status is alive with no injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESOLUTE ONYX RX
Common Device Name
STENT, CORONARY, DRUG-ELUTING
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway
MDR Report Key7395084
Report Number9612164-2018-00634
Device Sequence Number11164782
Product Code NIQ
UDI-Device Identifier00643169557185
UDI-Public00643169557185
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
P160043
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 04/04/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/09/2019
Device Catalogue NumberRONYX40022UX
Device Lot Number0008872002
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/09/2018
Initial Report FDA Received Date04/04/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-