| Catalog Number |
P4052 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
03/05/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
The physician was attempting to use a silverhawk directional atherectomy catheter to treat a plaque lesion with little calcification in the proximal sfa-popliteal artery (80% stenosis).The artery was non-tortuous.The device was prepped as per the ifu with no issues identified.The vessel was not pre or post dilated.During the initial advancement severe resistance was encountered.The blade of the device could not be released.The tip of the device embolised and was captured within the sheath.A new silverhawk was used to complete the procedure.No patient injury was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Product analysis: the silverhawk atk was returned for analysis.No other ancillary devices were included.The silverhawk was inspected and observed the cutter was within the cutter window of the distal assembly.Damage was noted on the cutter head.Biological material was observed within cutter window and on the cutter assembly.The cutter assembly was soaked in water.Under microscope it was discovered approximately 25% of the circumference of the distal rim of the cutter head was broken off.The piece which had broken off, was not returned with the silverhawk unit.The thumb-switch was pushed forward and the cutter impacted the distal rim of the cutter window (cutter crash).Event update: severe resistance was encountered during thumbswitch advancement.The thumbswitch was unable to open/ close the cutter.It was reported an unfamiliar sound was emitted from the motor and the motor experienced issues turning on and off.No intervention was required.Guidewire prolapse did not occur.The tip did not separate at the hinge pin.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|