Model Number TRUEMETRIX
Medical Device Problem Code High Test Results (2457)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 03/13/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Returned meter evaluated with no defect found.Test strips not returned for evaluation.Most likely underlying root cause: mlc-58-user had an inaccurate reference: self: the person is using themselves as the reference or how they feel at the time they run the blood test.Test strip udi# (b)(4).Note: manufacturer contacted customer (several attempts) in a follow-up call in order to ensure that the replacement products resolved the initial concern - unable to establish contact with the customer at this time.Product notification letter sent to customer to contact customer care.
 
Event or Problem Description
Consumer reported complaint for high blood glucose test results.The customer is concerned with tests results from results obtained of 153, 267, 295, 337, 318 mg/dl.The customer's expected fasting blood glucose test result range is 102 - 140 mg/dl.Customer stated truemetrix meter was reading high and then erratic after running the initial blood results.Customer stated she has had the truemetrix meter for over a year and this just started to happen within the last few weeks.The customer feels well and did not report any symptoms.Medical attention is not reported as a result of the actual blood glucose results.During the call on (b)(6) 2018, a back to back blood test was not performed by the customer.The product is stored according to specification in the dining room.The test strip lot manufacturer's expiration date is 01/14/2019 and open vial date is one week.The meter memory was reviewed for previous test result history: result 1:153mg/dl (b)(6) 8:28am fasting, result 2:267mg/dl (b)(6) 2:53pm fasting, result 3:295mg/dl (b)(6) 2:43pm fasting, result 4:337mg/dl (b)(6) 2:42pm fasting, result 5:318mg/dl (b)(6) 2:41pm fasting.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRUEMETRIX
Common Device Name
BLOOD GLUCOSE SYSTEM
Manufacturer (Section D)
TRIVIDIA HEALTH, INC.
2400 nw 55th court
fort lauderdale FL 33309
MDR Report Key7395623
Report Number1000113657-2018-00397
Device Sequence Number18849378
Product Code NBW
UDI-Device Identifier10612479210261
UDI-Public(01)10612479210261
Combination Product (Y/N)N
Initial Reporter StatePA
PMA/510(K) Number
K140100
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Remedial Action Replace
Type of Report Initial
Report Date (Section B) 04/04/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date01/14/2019
Device Model NumberTRUEMETRIX
Device Catalogue NumberRE4H01-81
Device Lot NumberMU2553
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/26/2018
Was the Report Sent to FDA? No
Event Location Home
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/13/2018
Initial Report FDA Received Date04/04/2018