| Model Number |
NX031Z |
| Medical Device Problem Code |
Loose or Intermittent Connection (1371)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Exemption number: e2014018.Manufacturing site evaluation: evaluation on-going.Device not returned.
|
| |
|
Event or Problem Description
|
|
Country of complaint: usa received via medwatch (mw5075167): tibia loosening femur of total knee, totally loose.Additional information has been requested.File will be updated when information is received.Med /watch submissions related to this report are: 9610612-2018-00159.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation: no product is at hand.Batch history review: the device quality and manufacturing history records have been checked for all lot numbers and found to be according to the specification, valid at the time of production.No similar incidents have been filed with products from the batches.Conclusion and root cause: no product available, and therefore a complete analysis is not possible, but based on the information available, it appears the root cause of the failure is not product related.There is the possibility that the root cause of the problem is most probably usage or patient related.Rational: according to the quality standard and dhr files, a material defect and production error can be excluded.Unfortunately due to a lack of data and without the product we can not determine the exact cause.Corrective action: product safety case 17-19.Any action regarding capa will be addressed with this case.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information - b5, d1, d2, d4 (model #, lot #, exp.Date, udi), d10, g4 (pma/510(k)), h4 this report is being submitted retrospectively per fda request.
|
| |
|
Event or Problem Description
|
|
Components in use listed as concomitant devices are: nx031z / as vega ps femoral comp.Cemented f5n r.Nx055z / as vega ps tibial plateau cemented t3.Nn264z / as tibial obturator d14mm.Nx130 / vega ps gliding surface t3/3+ 10mm.Nn483 / columbus patella 3-pegs p3 33x9mm.
|
| |
|
Search Alerts/Recalls
|