• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD ACCESS SYSTEMS HEMOSTAR LONG-TERM HEMODIALYSIS CATHETER 14.5F STANDARD KIT (STRAIGHT) (35CM); BLOOD ACCESS AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD ACCESS SYSTEMS HEMOSTAR LONG-TERM HEMODIALYSIS CATHETER 14.5F STANDARD KIT (STRAIGHT) (35CM); BLOOD ACCESS AND ACCESSORIES Back to Search Results
Model Number 5833350
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 03/13/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.The device has not been returned to the manufacturer for evaluation.A lot history review (lhr) of rebt1772 showed no other similar product complaint(s) from this lot number.
 
Event or Problem Description
It was reported by the nurse that during the use of the device for the third time after catheterization, it was found that the indentation at the thumb clip of the catheter could not be rebounded, resulting in narrow tubing which affects normal blood flow, then the venous pressure increased resulting in the device alarming.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEMOSTAR LONG-TERM HEMODIALYSIS CATHETER 14.5F STANDARD KIT (STRAIGHT) (35CM)
Common Device Name
BLOOD ACCESS AND ACCESSORIES
Manufacturer (Section D)
BARD ACCESS SYSTEMS
605 n. 5600 w.
salt lake city UT 84116
MDR Report Key7395995
Report Number3006260740-2018-00583
Device Sequence Number709650
Product Code MSD
UDI-Device Identifier00801741013362
UDI-Public(01)00801741013362
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K051748
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Nurse
Type of Report Initial
Report Date (Section B) 04/04/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number5833350
Device Catalogue Number5833350
Device Lot NumberREBT1772
Was Device Available for Evaluation? No
Event Location Hospital
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date04/04/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-