• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENDOLOGIX (TRIVASCULAR INC.) OVATION IX; ILIAC LIMB

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENDOLOGIX (TRIVASCULAR INC.) OVATION IX; ILIAC LIMB Back to Search Results
Model Number TV-IL1416100-J
Medical Device Problem Codes Failure To Adhere Or Bond (1031); Leak/Splash (1354)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 03/07/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Endologix will continue to investigate the reported event.The patient medical records and imaging studies will be requested for further evaluation by a clinical specialist.A final report will be submitted once the investigation of the reported event has concluded.
 
Event or Problem Description
An ovation ix iliac limb stent graft system was implanted to treat an abdominal aortic aneurysm.The patient returned for an unknown reason approximately 2 years post index procedure with a type ib endoelak of the right iliac limb.The physician is planning to re-intervene by extending the right iliac limb, however, a date for the re-intervention has not been scheduled.As of the date of this report, there have been no additional patient sequelae reported.
 
Additional Manufacturer Narrative
Based on the clinical assessment for this event, the most likely cause of the distal loss of seal was found to be unknown/undetermined.There was stent migration of the iliac limb, which likely contributed to the loss of seal, however, the cause of the migration could not be determined.There was no evidence of procedure related harms and the patient is reportedly being monitored.A review of the device quality records shows that the device demonstrated compliance to established procedures and specifications at the time of manufacture.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OVATION IX
Common Device Name
ILIAC LIMB
Manufacturer (Section D)
ENDOLOGIX (TRIVASCULAR INC.)
3910 brickway blvd.
santa rosa CA 95403
MDR Report Key7396061
Report Number3008011247-2018-00049
Device Sequence Number10775344
Product Code MIH
UDI-Device IdentifierM701TVIL1416100J1
UDI-Public+M701TVIL1416100J1/$$3160625FS05051507N
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
P120006
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/07/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date06/25/2016
Device Model NumberTV-IL1416100-J
Device Catalogue NumberTV-IL1416100-J
Device Lot NumberFS050515-07
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 03/07/2018
Supplement Date Received by Manufacturer03/07/2018
Initial Report FDA Received Date04/04/2018
Supplement Report FDA Received Date05/25/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
OVATION IX ILIAC LIMB: TV-IL1416120-J; OVATION IX MAIN BODY: TV-AB2680-J
Outcome Attributed to Adverse Event Other;
Patient Age72 YR
-
-