| Model Number |
MMT-723RNAL |
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
03/10/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
The customer reported via phone call the insulin pump had a keypad anomaly.The customer¿s blood glucose was 200 mg/dl at the time of incident.Customer stated that the insulin pump up and down buttons stopped working.No significant events leading to keypad anomaly were observed.Keypad anomaly troubleshooting was performed and confirmed that time is not advancing even after the battery has been changed.The customer was advised to discontinue use of pump and revert to back-up plan.The insulin pump is being replaced and is expected to return for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
No frozen screen or button error alarm noted.Device time or clock moved.Unit had intermittent buttons response due to flattened (creased) up and down buttons dome switch.No unlocked lcd keypad connector noted.
|
| |
|