| Model Number |
H7493952832400 |
| Medical Device Problem Code |
Bent (1059)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
03/07/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device is a combination product.(b)(4).Device evaluated by mfr.: the device has not been received for analysis.Upon receipt and completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental medwatch will be filed.(b)(4).
|
| |
|
Event or Problem Description
|
|
Same case as mdr id: 2134265-2018-02449.It was reported that stent damage occurred.The target lesion was located in the right coronary artery.After a 6f jr4 runway guide catheter was in place, a 4.00x12mm promus premier¿ drug-eluting stent was advanced but it was noted that the stent got caught with the guide catheter and was damaged.The device was removed and a 4.00x32mm promus premier¿ drug-eluting stent was then advanced; however, the stent got caught with the guide catheter and the physician felt that the second stent was damaged as well.The guide catheter was pulled out and a new guide catheter was then utilize to successfully deploy the stents.No patient complications were reported and the patient's status was great.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr: promus premier us mr 4.00 x 32mm stent delivery system was returned for analysis.A visual examination of the crimped stent identified damage.The mid-section of the stent was damaged and bunched causing the proximal end of the stent to move on the balloon 2mm distally.The distal end of the stent was still in the crimped position and was within max crimped stent profile measurement.The balloon cones were reviewed and no issues were noted.The balloon wings were tightly wrapped and evenly folded and were not subjected to positive pressure.A visual and tactile examination of the device revealed no issues with the hypotube.An examination of the shaft polymer extrusion revealed damage to the inner.The inner shaft polymer extrusion was bunched 3mm distal from the bi-component bond.The tip was visually examined and no issues were noted.No other issues were identified during the product analysis.The investigation conclusion is caused by other device as another device/drug/subsequent procedure caused the complaint event.(b)(4).
|
| |
|
Event or Problem Description
|
|
Same case as mdr id 2134265-2018-02449.It was reported that stent damage occurred.The target lesion was located in the right coronary artery.After a 6f jr4 runway guide catheter was in place, a 4.00x12mm promus premier drug-eluting stent was advanced but it was noted that the stent got caught with the guide catheter and was damaged.The device was removed and a 4.00x32mm promus premier drug-eluting stent was then advanced; however, the stent got caught with the guide catheter and the physician felt that the second stent was damaged as well.The guide catheter was pulled out and a new guide catheter was then utilize to successfully deploy the stents.No patient complications were reported and the patient's status was great.
|
| |
|