Investigation evaluation: a product evaluation was not performed in response to this report because the product said to be involved was not provided to cook for evaluation.The report could not be confirmed.The device history record for the lot number said to be involved was reviewed.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusion: we could not conduct a complete investigation because the product said to be involved was not returned for evaluation.A definitive cause for the reported observation could not be determined.The instructions for use (ifu) state, "if package is opened or damaged when received, do not use.Visually inspect with particular attention to kinks, bends and breaks.If an abnormality is detected that would prohibit proper working condition, do not use." the ifu states: "identify desired site based on previous findings from endoscopy, radiography and/or ct scans.Slowly introduce needle into accessory channel of endoscope and advance in short increments.Ensure needle is completely retracted and locked in place.Note: bends or kinks in needle caused by improper introduction may result in the inability to deploy fiducials." the user stated that the device was in an acute position and that the needle was bent at the distal end of the device as a result of the endoscope position.It is possible that if the needle is bent during introduction, becomes bent prior to reaching the identified site, or bends as a result of a hard mass, this could contribute to user's inability to deploy the fiducial markers.It is possible that if the needle is against or inside a hard mass while the user applies force and manipulates the directional controls of the endoscope, this could contribute to severe bending of the needle near the distal end.This contributes to advancement and/or retraction difficulties.Kinks in the needle can occur if the device experiences excessive pressure during product handling/preparation.Prior to distribution, all echotip ultra fiducial needles are subjected to a visual inspection to ensure device integrity.A review of the device history record confirmed that the lot said to be involved met all manufacturing requirements prior to shipment.Corrective action: corrective action is not warranted at this time based on the quality engineering risk assessment.Quality assurance will continue to monitor for complaint trends and reassess the risk assessment results as post market feedback continues to become available.
|