| Model Number |
TRUETRACK |
| Medical Device Problem Code |
Incorrect Or Inadequate Test Results (2456)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
02/16/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4) returned meter evaluated with no defect found.On qc investigation on 3/22/2018 - initial complaint was not confirmed.Additional failure found.Returned test strips produced high readings.Defect found.Most likely underlying root cause of high control results are due to improper storage: vial left open for an extended period of time.Test strip udi# (b)(4).Note: manufacturer contacted customer (several attempts) in a follow-up call in order to ensure that the replacement products resolved the initial concern - unable to establish contact with the customer at this time.Product notification letter sent to customer to contact customer care.
|
| |
|
Event or Problem Description
|
|
Consumer complaint of e-3 error; test strip error.E-3 error occurred upon insertion of test strip into meter port.Customer feels well and observed no symptoms.Medical intervention is not reported at the time of the call on (b)(6) 2018 as a result of the meter's readings.Blood test performed fasting during call on (b)(6) 2018 produced result of e-3 upon insertion of test strip into meter port.Verified storage of product is not within instructed specification since they are retained in the kitchen.Test strip lot manufacturer's expiration date is 12/20/2019 and open vial date is undisclosed.
|
| |
|