Investigation/evaluation: a review of the complaint history, device history record, documentation, drawing, instructions for use (ifu), manufacturing instructions, quality control, specifications, and visual inspection of the device was conducted during the investigation.One unopened c-aebs endobronchial blocker set was returned.A small, brown, unidentified piece of foreign matter was present in the sealed package.Foreign matter has been confirmed additionally, a document based investigation was performed.A review of the device history record for lot number 7853565 revealed no notable non-conformances related to the reported failure mode.A search of our complaint management system revealed this to be the only reported complaint from lot number 7853565.Each device is shipped with instructions for use (ifu) listing the indications for use, contraindications, warnings and precautions, and the correct deployment procedure.C_t_aebs_rev.3 [arndt endobronchial blocker set] states: "how supplied: supplied sterilized by ethylene oxide gas in peel-open packages.Do not use the product if there is doubt as to whether the product is sterile.Upon removal from package, inspect the product to ensure no damage has occurred." it is reasonable to suggest that foreign matter was introduced during the manufacturing process before the device's package was sealed.The cause of this is manufacturing related.This failure mode is being escalated per internal processes.We will continue to monitor for similar complaints and have notified the proper personnel about this event.Per the quality engineering risk assessment no further action is required.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|