MAUDE Adverse Event Report: COOK INC TRIPLE LUMEN POLYURETHANE CENTRAL VENOUS CATHETER SET; FOZ CATHETER, INTRAVASCULAR, THERAPEUTIC, SHORT-TERM LESS THAN 30 DAYS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK INC TRIPLE LUMEN POLYURETHANE CENTRAL VENOUS CATHETER SET; FOZ CATHETER, INTRAVASCULAR, THERAPEUTIC, SHORT-TERM LESS THAN 30 DAYS Back to Search Results
Model Number G47819
Medical Device Problem Code Device Contamination with Chemical or Other Material (2944)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 03/02/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).This report includes information known at this time.A follow up report will be submitted should additional relevant information become available.
 
Event or Problem Description
It was reported a hair was inside the primary sterile package of a triple lumen polyurethane central venous catheter set.This packaging anomaly was discovered during inspection.There were no adverse consequences to the patient as a result of this reported issue.The device did not make contact with the patient.
 
Additional Manufacturer Narrative
Investigation ¿ evaluation: a review of the complaint history, device history record, documentation, quality control, and specifications as well as a photographic inspection of the device was conducted during the investigation.The complaint device was not returned; therefore, no physical examinations could be performed.However, a document based investigation evaluation was performed.Pictures were provided and there is evidence to suggest/confirm that this device was manufactured out of specification.A review of the device history record shows no nonconforming events which could contribute to this failure mode.It should be noted there were no other reported complaints for this lot number.Based on the information provided and the results of our investigation, the root cause was determined to be related to the manufacturing process.We will continue to monitor for similar complaints and have notified the proper personnel about this event.Measures are being conducted to address this failure mode.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIPLE LUMEN POLYURETHANE CENTRAL VENOUS CATHETER SET
Common Device Name
FOZ CATHETER, INTRAVASCULAR, THERAPEUTIC, SHORT-TERM LESS THAN 30 DAYS
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
MDR Report Key7398087
Report Number1820334-2018-00862
Device Sequence Number16164062
Product Code FOZ
UDI-Device Identifier00827002478190
UDI-Public(01)00827002478190(17)201016(10)NS8296997
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K081113
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor,other,user facili
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/10/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberG47819
Device Catalogue NumberC-UTLM-701J-RSC-RD
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 03/13/2018
Supplement Date Received by Manufacturer04/18/2018
Initial Report FDA Received Date04/04/2018
Supplement Report FDA Received Date05/10/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-