STAAR SURGICAL COMPANY COLLAMER ULTRAVIOLET-ABSORBING POSTERIOR CHAMBER SINGLE PIECE FOLDABLE IOL; INTRAOCULAR LENS Back to Search Results
Model Number CC4204A
Medical Device Problem Code Positioning Failure (1158)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 03/01/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Work order search: no similar complaint type events were reported for units within the same lot.(b)(4).
 
Event or Problem Description
The reporter indicated that the surgeon attempted to implant a cc4204a, +23.0 diopter, intraocular lens in the patient's eye on (b)(6) 2018.It was reported that the plunger would not work as it would not advance.There was no patient contact with the device.Another staar lens was implanted.No other additional information was provided.
 
Additional Manufacturer Narrative
Device evaluation: product evaluation found no lens returned and two delivery systems.The first delivery system was found to have residue on product.Visual inspection found cartridge tip damaged and residue on device.The second delivery system was found returned in the lens box.Visual inspection found the plunger broken, the plunger tip damaged/stuck in cartridge and residue on the device.Claim# (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COLLAMER ULTRAVIOLET-ABSORBING POSTERIOR CHAMBER SINGLE PIECE FOLDABLE IOL
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
MDR Report Key7398186
Report Number2023826-2018-00493
Device Sequence Number16164059
Product Code HQL
UDI-Device Identifier00841542102452
UDI-Public00841542102452
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
P990013
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 03/06/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date12/31/2018
Device Model NumberCC4204A
Device Catalogue NumberN/A
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 03/06/2018
Supplement Date Received by Manufacturer05/03/2018
Initial Report FDA Received Date04/04/2018
Supplement Report FDA Received Date06/04/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-