| Model Number |
CQ2015A |
| Medical Device Problem Code |
Device Operates Differently Than Expected (2913)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
07/07/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product evaluation found that the lens was returned dry in a lens vial with clear surgical residue/debris on product.Visual inspection found residue on lens, the haptic torn and torn off from lens.Work order search: no similar complaint type events were reported for units within the same lot.(b)(4).
|
| |
|
Event or Problem Description
|
|
The reporter indicated that the surgeon attempted to implant a cq2015a, +20.5 diopter, intraocular lens in the patient's right eye (od) on (b)(6) 2017.During surgery, the haptic was torn off.There was no patient contact with the device.No other information was able to be provided.
|
| |
|