Model Number 60-01-00 |
Device Problems
Thermal Decomposition of Device (1071); Failure to Pump (1502)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 03/08/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
Patient identifier and weight not provided.This information will be provided in a supplemental report if made available.Livanova (b)(4) manufactures the centrifugal pump system with tubing clamp.The incident occurred in (b)(6).A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.Device not returned.
|
|
Event Description
|
Livanova (b)(4) received a report that a centrifugal pump system with tubing clamp pump did not start and after increasing the speed smoke came out of the engine during the procedure.There was no report of patient injury.
|
|
Manufacturer Narrative
|
The centrifugal pump system was returned to livanova (b)(4) for further investigation.During the evaluation it was found that a electronic component burned out on the board.During the investigation livanova (b)(4) determined that the burned out component was caused by incorrect user handling.
|
|