| Catalog Number |
THS-SX-C |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Codes |
Embolism (1829); Patient Problem/Medical Problem (2688)
|
| Date of Event |
03/07/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
The patient has a history of pad, renal failure.The physician intended to use the turbohawk and spider fx 5 mm and 6¿fr biotronik sheath to treat a severely calcified lesion, with 95-99% stenosis in the mid superficial femoral artery as per ifu.It was reported that severe calcification noted throughout the iliac, femoral and tibial vessels.The target lesions were crossed and predilated.The spider was placed in the distal popliteal artery.After pre-dilation the turbohawk was difficult to pass through and rotate, multiple passes were completed.Severe resistance was encountered removing the turbohawk, after fluoroscopic inspection it was observed that the wire was wrapped around the turbohawk.Numerous kinks were observed in the wire, attempts were made to untangle the wire from the turbohawk.The device and sheath were pulled back with resistance across aortic bifurcation, as a result the turbohawk nosecone separated at the hinge point from the catheter shaft in the common iliac artery.The spider wire did not fracture, the basket was in the proximal right sfa.Left sheath access was maintained the patient was transferred to hospital for a right femoral cutdown to retrieve the spider and nosecone.The or procedure was successful in removing the spider and nosecone of the turbohawk, with no complications reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The turbohawk btk directional atherectomy catheter was received for evaluation but the distal assembly and guidewire lumen were not included.No ancillary devices were received.A disc containing 12 dicom images folders was received for evaluation.The images are of the patient¿s right leg.Film analysis: evaluation of the film analysis showed contrast flow through the right common iliac artery, right femoral artery and right tibial artery.No component of the turbohawk, or ancillary devices were observed in the image of the right common iliac artery.In one image of the right tibial artery a guidewire navigated the full length of vessel is visible.In the images of the right femoral artery, signs of severe calcification in the mid superficial femoral artery were observed and a guidewire being navigated through the calcified vessel was also visible.One image showed treatment of the right femoral artery with a pta balloon.Device analysis: the device exhibited a tearing of the distal assembly just distal of the cutter window.The separation face exhibited torsional damage just distal of the cutter window and the distal edge of the cutter window exhibited signs of a cutter crash.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|