• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OBERDORF SYNTHES PRODUKTIONS GMBH AIMING ARM FOR RETROGRADE STANDARD LOCKING; GUIDE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OBERDORF SYNTHES PRODUKTIONS GMBH AIMING ARM FOR RETROGRADE STANDARD LOCKING; GUIDE Back to Search Results
Catalog Number 03.010.050
Medical Device Problem Code Failure to Align (2522)
Health Effect - Clinical Code No Code Available (3191)
Date of Event 03/13/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Patient¿s weight is unknown.Device is an instrument and is not implanted/explanted.(b)(4).The subject device has been received and the product evaluation is in progress.No conclusion can be drawn.A review of the device history records has been requested and is currently pending completion.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Device report from synthes europe reports an event in (b)(6) as follows: it was reported that a patient was implanted with a retrograde/antegrade femoral nail (rafn) to treat the distal femur fracture on (b)(6) 2018.During the procedure the two (2) distal screws missed the nail hole (rafn) which caused the screws inserted posteriorly.The connecting screw was tightly secured to the insertion handle and pre-checking of the nail and jig alignment was done and it was aligned perfectly.Surgeon doesn¿t know why the jig would have missed the nail, given that the connecting screw was tight too.The surgeon had to remove the 2 screws and replace it with a spiral blade for the nail.There was a three (3) minute surgical delay to the procedure.No patient consequences reported.This report is for one (1) aiming arm for retrograde standard locking.This is report 2 of 3 for complaint (b)(4).
 
Additional Manufacturer Narrative
Dhr review was completed.Part number: 03.010.050.Synthes lot number: 9449870.Release to warehouse date: 26.May.2015.Manufacturing site: (b)(4).No ncrs were generated during production.Review of the device history record(s) showed that there were no issues during the manufacture of the product that would contribute to this complaint condition.H3, h6: customer quality conducted an investigation of the returned device.The returned aim-arm f/expert r/afn retrogr was forwarded to the manufacturing site and was checked for conformance to the specifications.Received 1 article of aim-arm f/expert r/afn retrogr for manufacturing investigation.The article is in an used condition.Traces of use are clearly visible on the two locating pins.The review of the device history record revealed that this instrument was manufactured in may 2015 according to the specifications.No ncrs were generated during production.Review of the device history record showed that there were no issues during the manufacturing of the product.During manufacturing investigation, all relevant and significant characteristics like dimensions were checked and additionally, the article passed all functional tests.All checked characteristics are within the specifications.In accordance to the test results of 03.010.050 aim-arm f/expert r/afn no misalignment or any other production error has been found.The only reference to the reported issue was detected during visual inspection.Clearly visible traces are present on the locating pins, this can be a sign of forcing, after the aim-arm f/expert r/afn was mounted on the insert-handle long f-expert fn 03.010.046.The functionality is not affected by the traces on the location pins.No manufacturing related issue was identified and/or confirmed, therefore review to the specific prm and prm line is not applicable.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AIMING ARM FOR RETROGRADE STANDARD LOCKING
Common Device Name
GUIDE
Manufacturer (Section D)
OBERDORF SYNTHES PRODUKTIONS GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
MDR Report Key7399644
Report Number8030965-2018-52795
Device Sequence Number11197774
Product Code FZX
UDI-Device Identifier07611819784093
UDI-Public(01)07611819784093(10)9449870
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date (Section B) 03/13/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number03.010.050
Device Lot Number9449870
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/26/2018
Initial Date Received by Manufacturer 03/13/2018
Supplement Date Received by Manufacturer05/17/2018
Initial Report FDA Received Date04/05/2018
Supplement Report FDA Received Date05/22/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age38 YR
-
-