Patient¿s weight is unknown.Device is an instrument and is not implanted/explanted.(b)(4).The subject device has been received and the product evaluation is in progress.No conclusion can be drawn.A review of the device history records has been requested and is currently pending completion.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Dhr review was completed.Part number: 03.010.050.Synthes lot number: 9449870.Release to warehouse date: 26.May.2015.Manufacturing site: (b)(4).No ncrs were generated during production.Review of the device history record(s) showed that there were no issues during the manufacture of the product that would contribute to this complaint condition.H3, h6: customer quality conducted an investigation of the returned device.The returned aim-arm f/expert r/afn retrogr was forwarded to the manufacturing site and was checked for conformance to the specifications.Received 1 article of aim-arm f/expert r/afn retrogr for manufacturing investigation.The article is in an used condition.Traces of use are clearly visible on the two locating pins.The review of the device history record revealed that this instrument was manufactured in may 2015 according to the specifications.No ncrs were generated during production.Review of the device history record showed that there were no issues during the manufacturing of the product.During manufacturing investigation, all relevant and significant characteristics like dimensions were checked and additionally, the article passed all functional tests.All checked characteristics are within the specifications.In accordance to the test results of 03.010.050 aim-arm f/expert r/afn no misalignment or any other production error has been found.The only reference to the reported issue was detected during visual inspection.Clearly visible traces are present on the locating pins, this can be a sign of forcing, after the aim-arm f/expert r/afn was mounted on the insert-handle long f-expert fn 03.010.046.The functionality is not affected by the traces on the location pins.No manufacturing related issue was identified and/or confirmed, therefore review to the specific prm and prm line is not applicable.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|