• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PENUMBRA, INC. BENCHMARK DELIVERY CATHETER; DQY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PENUMBRA, INC. BENCHMARK DELIVERY CATHETER; DQY Back to Search Results
Catalog Number BMK6F105M
Medical Device Problem Codes Kinked (1339); Device Damaged Prior to Use (2284); Out-Of-Box Failure (2311)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 03/06/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Results: the benchmark was kinked in multiple locations along its length.Conclusions: evaluation of the returned benchmark revealed kinks along its length.If the catheter was removed from its packaging tube at extreme angles, kinks may result.Based on the returned condition, several kinks on the device were likely incidental and may have occurred during packaging for return to penumbra.Penumbra catheters are inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
 
Event or Problem Description
During preparation for a medical procedure, the hospital staff found that the benchmark 6f 071 delivery catheter (benchmark) was kinked upon removal from the packaging at the back table.The damage to the benchmark was found prior to use.Therefore, the benchmark was not used in the procedure.The procedure was completed using a new benchmark.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BENCHMARK DELIVERY CATHETER
Common Device Name
DQY
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key7399707
Report Number3005168196-2018-00695
Device Sequence Number19200175
Product Code DQY
UDI-Device Identifier00814548014043
UDI-Public00814548014043
Combination Product (Y/N)N
Initial Reporter CountryHK
PMA/510(K) Number
K142321
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor,foreign,health pr
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 03/06/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Radiologic Technologist
Device Expiration Date11/09/2020
Device Catalogue NumberBMK6F105M
Device Lot NumberF80504
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/02/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/06/2018
Initial Report FDA Received Date04/05/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-