Model Number ZCT300
Medical Device Problem Code Unintended Movement (3026)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 12/01/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If explanted; give date: not applicable as the lens remains implanted.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported that a tecnis toric intraocular lens (iol) was placed at 008 degrees in the patient¿s right eye during surgery on (b)(6) 2017.The placement was measured at the 6-month visit on (b)(6) 2017 and was found to be at 178 degrees.It was noted that the implant procedure was completed successfully.The lens remains implanted.The patient is doing well and the outcome is not interfering with daily activities.There was no patient complication and/or related injury and there are no iol repositioning is planned.No additional information provided.
 
Additional Manufacturer Narrative
Device evaluation: the product testing could not be performed as the product was not returned.The reported complaint cannot be confirmed.Manufacturing record review: the manufacturing process record was evaluated and no deviation was found during process related to the complaint issue reported.There was no discrepancy found during the mrr (manufacturing record review).The product was manufactured and released according to the specifications.A search revealed that no similar complaint was received from this production order.Labeling review: the directions for use (dfu) was reviewed.The dfu adequately provides instructions and precautions for the proper use and handling of the product.Conclusion: based on the investigation results there is no indication of a product quality deficiency.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS
Common Device Name
TORIC IOLS
Manufacturer (Section D)
JOHNSON & JOHNSON SURGICAL VISION, INC.
santa ana CA 92705
MDR Report Key7399910
Report Number9614546-2018-00316
Device Sequence Number16449261
Product Code HQL
UDI-Device Identifier05050474553286
UDI-Public(01)05050474553286(17)180207
Combination Product (Y/N)N
Initial Reporter StateMD
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/31/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Expiration Date02/07/2018
Device Model NumberZCT300
Device Catalogue NumberZCT300U180
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 03/06/2018
Supplement Date Received by Manufacturer05/02/2018
Initial Report FDA Received Date04/05/2018
Supplement Report FDA Received Date05/31/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age77 YR
-
-