• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT MANUFACTURING INC CLINICAL CHEMISTRY TOTAL BILIRUBIN; TOTAL BILIRUBIN REAGENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MANUFACTURING INC CLINICAL CHEMISTRY TOTAL BILIRUBIN; TOTAL BILIRUBIN REAGENT Back to Search Results
Catalog Number 06L45-41
Medical Device Problem Code Incorrect Or Inadequate Test Results (2456)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 02/21/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Evaluation of the customer issue included a review of the complaint text, a search for similar complaints, device history review, instrument log review, and labeling review.No adverse trend was identified for the customer issue.Device history review did not identify any issues that may have caused the customer issue.Labeling was reviewed and found to be adequate.Log review did not identify any issues that contributed to the customer issue.Based on all available information and abbott diagnostics evaluation, no systemic issue was identified and no product deficiency was found.This issue was previously reported under mdr number 1415939-2018-00045 under a different suspect device manufacturer.
 
Event or Problem Description
The customer observed multiple falsely elevated total bilirubin results while using the clinical chemistry total bilirubin reagents.The following data was provided.Sid (b)(6) initial 1.636 mg/dl, repeat 0.509 mg/dl, sid (b)(6) initial 1.807 mg/dl, repeat 0.717 mg/dl, sid (b)(6) initial 1.646 mg/dl, repeat 0.493 mg/dl, sid (b)(6) initial 1.795 mg/dl, repeat 0.674 mg/dl, sid (b)(6) initial 1.582 mg/dl, repeat 0.443 mg/dl, sid (b)(6) initial 1.581 mg/dl, repeat 0.431 mg/dl, sid (b)(6) initial 2.088 mg/dl, repeat 1.045 mg/dl, sid (b)(6) initial 1.711 mg/dl, repeat 0.578 mg/dl, sid (b)(6) initial 1.654 mg/dl, repeat 0.512 mg/dl, sid (b)(6) initial 1.778 mg/dl, repeat 0.66 mg/dl, sid (b)(6) initial 1.835 mg/dl, repeat 0.781 mg/dl, sid (b)(6) initial 1.648 mg/dl, repeat 0.528 mg/dl, sid (b)(6) initial 1.623 mg/dl, repeat 0.453 mg/dl, sid (b)(6) initial 1.605 mg/dl, repeat 0.472 mg/dl, sid (b)(6) initial 1.474 mg/dl, repeat 0.34 mg/dl, sid (b)(6) initial 1.699 mg/dl, repeat 0.592 mg/dl, sid (b)(6) initial 1.548 mg/dl, repeat 0.427 mg/dl, sid (b)(6) initial 1.734 mg/dl, repeat 0.608 mg/dl, sid (b)(6) initial 1.425 mg/dl, repeat 0.278 mg/dl, sid (b)(6) initial 1.756 mg/dl, repeat 0.625 mg/dl, sid (b)(6) initial 1.648 mg/dl, repeat 0.505 mg/dl, sid (b)(6) initial 1.373 mg/dl, repeat 0.203 mg/dl, sid (b)(6) initial 1.412 mg/dl, repeat 0.262 mg/dl, sid (b)(6) initial 1.335 mg/dl, repeat 0.193 mg/dl, sid (b)(6) initial 1.915 mg/dl, repeat 0.828 mg/dl, sid (b)(6) initial 1.383 mg/dl, repeat 0.219 mg/dl, sid (b)(6) initial 1.909 mg/dl, repeat 0.802 mg/dl, sid (b)(6) initial 2.137 mg/dl, repeat 1.087 mg/dl, sid (b)(6) initial 1.603 mg/dl, repeat 0.469 mg/dl, sid (b)(6) initial 1.695 mg/dl, repeat 0.564 mg/dl, sid (b)(6) initial 1.508 mg/dl, repeat 0.337 mg/dl, sid (b)(6) initial 1.896 mg/dl, repeat 0.804 mg/dl, sid (b)(6) initial 1.638 mg/dl, repeat 0.514 mg/dl, sid (b)(6) initial 1.494 mg/dl, repeat 0.321 mg/dl, sid (b)(6) initial 2.027 mg/dl, repeat 0.972 mg/dl, sid (b)(6) initial 1.705 mg/dl, repeat 0.594 mg/dl, sid (b)(6) initial 1.658 mg/dl, repeat 0.53 mg/dl, sid (b)(6) initial 1.674 mg/dl, repeat 0.564 mg/dl, sid (b)(6) initial 1.738 mg/dl, repeat 0.608 mg/dl, sid (b)(6) initial 1.615 mg/dl, repeat 0.481 mg/dl, sid (b)(6) initial 1.695 mg/dl, repeat 0.557 mg/dl, sid (b)(6) initial 1.668 mg/dl, repeat 0.54 mg/dl, sid (b)(6) initial 1.689 mg/dl, repeat 0.568 mg/dl, sid (b)(6) initial 1.394 mg/dl, repeat 0.233 mg/dl, sid (b)(6) initial 2.092 mg/dl, repeat 1 mg/dl, sid (b)(6) initial 1.953 mg/dl, repeat 0.87 mg/dl, sid (b)(6) initial 1.471 mg/dl, repeat 0.288 mg/dl, sid (b)(6) initial 1.778 mg/dl, repeat 0.679 mg/dl, sid (b)(6) initial 1.742 mg/dl, repeat 0.627 mg/dl, sid (b)(6) initial 1.782 mg/dl, repeat 0.658 mg/dl, sid (b)(6) initial 2.909 mg/dl, repeat 1.916 mg/dl, sid (b)(6) initial 1.44 mg/dl, repeat 0.288 mg/dl, sid (b)(6) initial 1.752 mg/dl, repeat 0.623 mg/dl, sid (b)(6) initial 1.709 mg/dl, repeat 0.571 mg/dl, sid (b)(6) initial 1.957 mg/dl, repeat 0.867 mg/dl, sid (b)(6) initial 3.045 mg/dl, repeat 2.082 mg/dl, sid (b)(6) initial 1.959 mg/dl, repeat 0.854 mg/dl, sid (b)(6) initial 1.636 mg/dl, repeat 0.493 mg/dl, sid (b)(6) initial 1.825 mg/dl, repeat 0.733 mg/dl, sid (b)(6) initial 2.491 mg/dl, repeat 1.467 mg/dl, sid (b)(6) initial 1.45 mg/dl, repeat 0.287 mg/dl, sid (b)(6) initial 2.163 mg/dl, repeat 1.122 mg/dl no impact to patient management was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLINICAL CHEMISTRY TOTAL BILIRUBIN
Common Device Name
TOTAL BILIRUBIN REAGENT
Manufacturer (Section D)
ABBOTT MANUFACTURING INC
1921 hurd drive
irving TX 75038
MDR Report Key7400545
Report Number1628664-2018-00136
Device Sequence Number1470816
Product Code CIG
UDI-Device Identifier00380740011383
UDI-Public00380740011383
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
K060574
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Health Professional
Type of Report Initial
Report Date (Section B) 04/05/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/28/2018
Device Catalogue Number06L45-41
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/02/2018
Initial Report FDA Received Date04/05/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-