MAUDE Adverse Event Report: MEDTRONIC NEUROSURGERY CSF VENTRICULAR PORT, 16 MM, SIDE INLET; PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVENTRICULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NEUROSURGERY CSF VENTRICULAR PORT, 16 MM, SIDE INLET; PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVENTRICULAR Back to Search Results
Model Number 44010
Medical Device Problem Code Appropriate Device Problem Term/Code Not Available (3191)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 04/01/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that the surgeon was performing a tumor resection for the patient, and during the operation, they found that the device couldn't drain cerebrospinal fluid normally.It was stated that the surgeon immediately replaced the device with another one.The patient's status at the time of the report was alive-no injury.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Additional information received reported that the doctor confirmed that the product was unable to drain cerebrospinal fluid properly.The doctor thought the product quality problem led to the cerebrospinal fluid not being able to outflow.
 
Additional Manufacturer Narrative
The returned port met the requirements for leak testing.Proteinaceous debris was observed within the interior and exterior of the valve.Approximately 23 cm of the returned catheter was patent and met the requirements for leak testing.All ports and catheters are 100% inspected at the time of manufacture.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CSF VENTRICULAR PORT, 16 MM, SIDE INLET
Common Device Name
PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVENTRICULAR
Manufacturer (Section D)
MEDTRONIC NEUROSURGERY
125 cremona drive
goleta CA 93117
MDR Report Key7400719
Report Number2021898-2018-00182
Device Sequence Number19200203
Product Code LKG
UDI-Device Identifier00643169497108
UDI-Public00643169497108
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K874468
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2018
Reporter Type Manufacturer
Report Source foreign,health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 05/15/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/31/2021
Device Model Number44010
Device Catalogue Number44010
Device Lot NumberE22922
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/16/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/02/2018
Supplement Date Received by Manufacturer04/22/2018
05/15/2018
Initial Report FDA Received Date04/05/2018
Supplement Report FDA Received Date05/01/2018
05/15/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age66 YR
Patient Weight58
-
-