MAUDE Adverse Event Report: MEDTRONIC MINIMED PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-723RNAH; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC MINIMED PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-723RNAH; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR Back to Search Results
Model Number MMT-723RNAH
Health Effect - Clinical Code Hyperglycemia (1905)
Date of Event 09/15/2017
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The insulin pump was received with missing segment/partial display, problem isolated to lcd board.Unable to perform operating currents, self-test, unexpected restart error test, rewind test, basic occlusion test, occlusion test, prime test, excessive no delivery test and displacement test due to missing segment.The insulin pump had severely scratched lcd window, cracked battery tube threads, cracked reservoir tube lip, broken belt clip slot, scratched reservoir tube window, stained address/serial number label, stained end cap sticker and corroded battery tube.
 
Event or Problem Description
The customer reported via phone call that they were hospitalized due to diabetic ketoacidosis on (b)(6) 2017 with blood glucose of 700 mg/dl.Customer reported insulin pump was not working, crack on insulin pump and lcd screen not visible.The customer experienced symptoms such as vomiting, diarrhea.The customer was treated with shots.Customer reports they have contacted their healthcare professional regarding the high blood glucose.The customer was not wearing the insulin pump during the hospitalization.Customer was off the device prior to hospitalization for greater than 48 hours.Customer declined trouble shooting because it was a past event and hasn't been on the pump.Customer stated that he went into the hospital he was in for 4 weeks then let go.Customer stated when he was let go he wasn't on the pump and has been on shots the whole remainder of time when he was put back in the hospital.Customer stated he doesn't know what his blood sugar was the first time going into the hospital but the second time he went in his blood sugar was 700.Customer declined trouble shooting for the no delivery.The insulin pump was returned for analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-723RNAH
Common Device Name
PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR
Manufacturer (Section D)
MEDTRONIC MINIMED
18000 devonshire st.
northridge CA 91325 1219
MDR Report Key7401175
Report Number2032227-2018-02359
Device Sequence Number13630684
Product Code OYC
UDI-Device Identifier00613994748720
UDI-Public(01)00613994748720
Combination Product (Y/N)N
Initial Reporter StateIA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Patient
Type of Report Initial
Report Date (Section B) 04/05/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberMMT-723RNAH
Device Catalogue NumberMMT-723RNAH
Device Lot NumberA000168687
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/15/2017
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/26/2017
Initial Report FDA Received Date04/05/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age56 YR
Patient Weight205
-
-