• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AV-TEMECULA-CT ARMADA 18 PTA CATHETER; PERIPHERAL DILATATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AV-TEMECULA-CT ARMADA 18 PTA CATHETER; PERIPHERAL DILATATION CATHETER Back to Search Results
Catalog Number 1013470-040
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 03/15/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Internal file number - (b)(4): during processing of this complaint, attempts were made to obtain complete event, patient and device information.Evaluation summary: visual and functional inspection was performed on the returned device.The hub leak was confirmed.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history of the reported lot revealed no other similar incidents.The investigation determined the reported hub leak appears to be related to a potential product quality issue.The issue is being addressed per internal operating procedures.Abbott vascular will continue to trend the performance of these devices.
 
Event or Problem Description
It was reported that during preparation of the 6.0/40 mm armada 18 balloon catheter, saline was observed leaking from the hub of the device.The device was not used on the patient.A new armada 18 balloon catheter was used successfully for the case.There was no clinically significant delay in the procedure reported.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARMADA 18 PTA CATHETER
Common Device Name
PERIPHERAL DILATATION CATHETER
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
MDR Report Key7401217
Report Number2024168-2018-02512
Device Sequence Number952725
Product Code LIT
UDI-Device Identifier08717648207808
UDI-Public08717648207808
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
K151317
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial
Report Date (Section B) 04/05/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/30/2020
Device Catalogue Number1013470-040
Device Lot Number7013141
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/23/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/15/2018
Initial Report FDA Received Date04/05/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-