MAUDE Adverse Event Report: PENUMBRA, INC. RUBY COIL; HCG, KRD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PENUMBRA, INC. RUBY COIL; HCG, KRD Back to Search Results
Catalog Number RBY4C0204
Medical Device Problem Codes Detachment Of Device Component (1104); Difficult or Delayed Activation (2577); Physical Resistance (2578)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 03/08/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The product was not returned for evaluation.Without the return of the device, the root cause of the problem cannot be determined.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.This report is associated with mfr report number: 3005168196-2018-00715.
 
Event or Problem Description
The patient was undergoing a coil embolization procedure in the left hepatic artery using ruby coils and a ruby coil detachment handle (handle).During the procedure, the physician successfully placed multiple coils using a non-penumbra microcatheter.The physician was then unable to detach a ruby coil using the handle.Therefore, the physician attempted to manually detach the ruby coil, however, the ruby coil still did not detach.The pusher assembly was, therefore, retracted to be removed, however, the physician felt resistance while retracting and the ruby coil unintentionally detached in the target location.The procedure was completed using an additional ruby coil, the same microcatheter, and the same handle.There was no report of an adverse effect to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RUBY COIL
Common Device Name
HCG, KRD
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key7401473
Report Number3005168196-2018-00714
Device Sequence Number1393058
Product Code HCG
UDI-Device Identifier00814548013138
UDI-Public00814548013138
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K120330
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 03/08/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Expiration Date11/08/2023
Device Catalogue NumberRBY4C0204
Device Lot NumberF66393
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/08/2018
Initial Report FDA Received Date04/05/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age60 YR
Patient Weight80
-
-